iSOLV TUF336 Ultrasonic Clamp-On Flowmeter

The TUF336 is an ultrasonic permanent/fixed clamp-on flow metering solution for measuring flow rate and total flow with a volume pulse output and optional Modbus or 4-20mA flow proportional output, which can be used as a stand alone meter or as part of an integral management system.
Simple to install – connect power and enter the pipe inside diameter, adjust the sensors and clamp-on the pipe -no specialist skills or tools required!
TUF336 clamp-on ultrasonic flow meter is a cost effective alternative to traditional in-line meter installation, plus dry servicing, providing minimum downtime and maximum availability.

ENQUIRY
Request Price, Order and Brochure click here.
Features


• Factory configured – minimal site configuration
• Adjustable guide rail & sensor assembly – simple installation
• Clamp-on – non-invasive & low install cost
• LCD with backlight – install & service information
• Integral pulse or optional modbus and 4-20mA outputs – aM&T & BEM’s compatible
• Adjustable guide rail & sensor assembly for simple installation
• Non-invasive solution
• Low installation cost
• LCD with backlight
• Transmitter can be placed both locally and remotely
• Economical version catering to small pipe size (up to 180mm)

Application


• Water/wastewater
• Building Energy Management
• Semi Conductor Industry
• Pharmaceutical
• Petrochemical
• Food and beverage

Specifications


• Measurement Technique:
  - Ultrasonic transit time method for flow measurement.
  - Turn Down Ratio: 200:1
• Accuracy: +/- 1% – 3% of flow reading for >0.3m/s (1 ft/s)
• Flow Velocity Range: 0.1m/s – 10m/s (0.3 ft/s – 32 ft/s)
• Pipe Range: Available in 2 options:
  - 22mm to 115mm OD
  - 125mm to 180mm OD
  - Note Pipe size is dependent on pipe material and internal diameter
• Water Temp Range: 0°C – 85°C (32°F – 185°F)
• Pulse Output:
  - Pulse or Frequency
  - Pulse for Volume flow and Alarms
  - Frequency for flow rate
  - The pulse output can be configured as a loss of signal or low flow alarm
• 4-20mA Output: Optional 4-20mA flow proportional output
• Modbus Communication: Optional Modbus RTU slave, RS485 serial link hardware layer. Modbus connection cable is 1m.
• External Power Supply: 12V – 24V +/- 10% AC/DC at 7 watts per unit. Optional plug in 12V power supply
• Electronics Enclosure: IP54
• Input/Output Cable: 5m x 6 core for power in, 4-20mA and pulse out
• Dimensions: 250mm x 48mm x 90mm (10” x 2” x 4”) (electronics + guide rail)

*Note: Specifications are subject to change without notice